CONFIGURATION(cont.) StepFive
HowtoconfiguretheOperationalrangeofEchoSwitch®
Before starting, take two measurements.1) Measure the
distancefromthebottomofthesensor (transducerface)tothe
bottomofthetank.ThiswillbetheHEIGHTsetting.2)Measure
thedistancefromthebottomofthetanktotheMaximumLevel
ofLiquid(Full).ThiswillbetheFILL‐Hsetting.
1. InTOP‐LEVELMENUmode,selectTANK.
2. SelectHEIGHT.
3. Using the UP and DOWN buttons, set the HEIGHT of the
tank(transducerfacetothebottomofthetank).
4. To enter the value, press and hold SELECT for 2 seconds
andrelease.SAVEDwilldisplay.
5. WhenFILL‐Happears,pressSELECT.
6. UsingtheUPandDOWNbuttons,setthe FillHeight(FILL‐
H) of the tank (max. level of liquid to the bottom of the
tank).
7. To enter the value, press and hold SELECT for 2 seconds
andrelease.SAVEDwilldisplay.
8. SelectEXITtoreturntoTOP‐LEVELMENU.
Note:The Height and Fill‐Height (Fill‐H) settings also determine the 4 to 20 mA current span.The Height
settingdeterminesthe4mApositionandtheFill‐Hsettingdeterminesthe20mAposition.
Note:Tospeedupthescrollingofthevaluesonthedisplay,holddowntheSELECTbuttonwhileholdingdown
theUPorDOWNbuttons.